Karl Fischer unit
Submitted by Anonymous on Tue, 2016-07-12 15:02
Titration Method Should be Coulometric Karl Fischer TitrationElectrolysis Control : Only AutomaticEnd Point Detection : AC polarizationMeasuring range : Possible 1μg - 200mg waterMoisture range: 1ppm - 100% waterMax. Sensitivity: should be minimum 0.1μgMax. titration speed: At least 2.5 mg per minute or BetterMax. current: 400 maDrift compensation: should be automatically controlled only Precision: 10-100μg ± 4μg, 100μg-1mg ± 4μg, above 1mg ±0.4% or Better